Bionime Blood Glucose Test Strip GS100 50 PSC ( 2 vials of 25 )  

Big Single Button, Easy to Get Started
Our mission is to ensure the accuracy of blood glucose measurements, and user-friendly design allows you to operate more simple and convenient.

Proprietary Rightest® strip

Proprietary Rightest™ strip
With Validus Technology™ strip design, Rightest™ strips ensure high accuracy and precision in readings.

Blood contact prevention

Blood contact prevention
Larger, thicker test strips help with dexterity issues. The innovate side-insert design ensures better hygiene.

Quick result

Quick result
Accurate measurements in 8 seconds.

Tiny blood sample required

Tiny blood sample required
Only 1.4 microliters (0.0014c.c.) of blood sample required.

GM100 Specifications
Test strip GS100
Coding Coding is not required
Enzyme / Measurement technology GOD / Electrochemical sensor
Blood sample source Capillary whole blood
Minimum sample volume 1.4 microliters  (0.0014c.c.)
HCT range 35 - 48%
Measuring range 10 - 600mg/dL (0.6 - 33.3 mmol/L)
Meter dimensions (L x W x H) 95.0 x 43.8 x 13.0 mm (3.74 x 1.72 x 0.51 inch)
Display area (L x W) 39.0 x 29.0 mm (1.49 x 1.14 inch)
Weight (with battery) 43.0g
Power supply 1 CR2032 battery
Memory capacity 150 blood glucose test results with date and time
Operating temperature 10 - 40°C (50 - 104°F)
Operating relative humidity 10 - 90%
Meter storage conditions -10 - 60°C (14 - 140°F)
Test strip storage conditions 4 - 30°C (39 - 86°F), < 90% relative humidity
Accuracy 95% of the measured glucose values shall fall within either
Glucose concentrations < 5.5mmol/L (100mg/dL), difference within ± 0.83mmol/L (15mg/dL)
Glucose concentrations ≥ 5.5mmol/L (100mg/dL), bias within ± 15%
Precision Glucose concentrations < 5.5mmol/L (100mg/dL), SD < 0.28mmol/L (5mg/dL)
Glucose concentrations ≥ 5.5mmol/L (100mg/dL), CV < 5%
Fully comply with ISO 15197:2013 (For other received credentials, please contact us.)

 